EEG functionality
- Number of channels: (up to) 32
- Sampling rate: 500 SPS
- Bandwidth: 0 to 125 Hz (DC coupled)
- Resolution: 24 bits – 0,05 µV resolution
- Noise: < 1 µV RMS
- Common mode rejection ratio: -115 dB
- Input impedance: 1 GΩ
Stimulation functionality
- Number of channels: (up to) 32
- Sampling rate: 1000 SPS
- Stimulation types: tDCS, tACS and tRNS
- Customized waveform
- Frequency range: 0-250 Hz (tACS); 0-500 Hz (tRNS)
- Configurable bandpass filter for tRNS
- Maximum current per channel: ± 4mA
- Current accuracy: 1%
- Current resolution: 1 μA
- Current is configurable independently for each channel
- Configurable Ramp-up and Ramp-down times
- Voltage: ± 15 V per electrode (30 V potential difference)
- Sham and Double blind modes available
Stimulation safety:
- Max current per electrode: 2 mA
- Max total injected current (by all electrodes, any time): 4 mA
- Max duration per session: 1 hour
- Stimulation session must be pre-programmed
- Impedance check before and during stimulation
- Abort functionality: possible at anytime
NECBOX (Neuroelectrics Control Box)
- Battery: USB rechargeable Li-Ion battery
- Battery operating time: 8 hours (combined EEG/tCS)
- Data transmisstion: wireless (wi-fi network) or wired (USB cable)
- microSD card slot (for offline data storage)
- Output: EDF+, ASCII data files or TCP/IP raw data streaming
- Accelerometer data recording: tri-axial @ 100 SPS